Medical Radiology and Radiation Safety. 2025. Vol. 70. № 3
DOI:10.33266/1024-6177-2025-70-3-108-116
D.V. Arefyeva, V.B. Firsanov, S.V. Yarmiychuk, A.V. Petushok
Application of the Monte-Carlo Method
for Calibration of a Gamma-ray Scintillation Spectrometer
Scientific Research Institute of Industrial and Marine Medicine, St. Petersburg, Russia
Contact person: D.V. Arefyeva, e-mail: This email address is being protected from spambots. You need JavaScript enabled to view it.
ABSTRACT
Purpose: To develop a method for calibration of a gamma-ray scintillation spectrometer using the Monte Carlo method.
Material and methods: The subject of the study was a gamma-ray spectrometer designed to measure the energy distribution (spectrum) and determine the activity of gamma-emitting radionuclides. Experimental studies were carried out with a set of exemplary measures of special-purpose activity with radionuclides 241Am, 152Eu, 60Co and 137Cs uniformly deposited on an ion exchange resin. Calibration of the spectrometer was carried out using the MCC 3D program (Monte Carlo Calculations 3D), modeling of the hardware spectrum was performed using the MCA program (MultiChannel Analyzer).
Results: The comparison of experimental and simulated spectra was carried out in the following energy intervals: the interval corresponding to the total peak of total absorption (PTA) for gamma energy lines 1173.2 keV and 1332.5 keV for 60Co and PTA for gamma energy line 661.7 keV for 137Cs; intervals corresponding to Compton scattering in the angle range (30–60)°, (60–90)° and (90–180)° (for the 60Co, the average gamma radiation energy of 1252.9 keV was considered); the interval corresponding to multiple scattering with an energy above 100 keV. It was found that the largest deviation of the simulated spectrum from the experimental one is 12 % for the interval corresponding to multiple scattering, which indicates the possibility of spectrum identity. This assumption was verified for each energy interval using the Pearson consensus criterion. A maximum value of χ2 equal to 6.6 was obtained for the energy interval corresponding to Compton scattering in the angle range (60–90)°, which indicates the acceptability of the hypothesis of the identity of the experimental and simulated spectra. Validation of the proposed method showed that the discrepancy between the calculated and passport activity of the sample was no more than 13 %, which indicates the possibility of using the method for calibration of the gamma spectrometer. The dependences of the efficiency of registration in the PTA on the density of the counting sample are calculated using simulated hardware spectra of single activity.
Conclusion: The proposed method makes it possible to calibrate the spectrometer to calculate the specific activity in samples at various densities and energies using spectrometric equipment equipped with inorganic scintillation crystals.
Keywords: gamma-ray spectrometer, Monte Carlo method, calibration, radiation safety
For citation: Arefyeva DV, Firsanov VB, Yarmiychuk SV, Petushok AV. Application of the Monte-Carlo Method for Calibration of a Gamma-ray Scintillation Spectrometer. Medical Radiology and Radiation Safety. 2025;70(3):108–116. (In Russian). DOI:10.33266/1024-6177-2025-70-3-108-116
References
1. Monte Carlo N-Particle Transport Code. URL: https://ru.wikipedia.org/wiki/MCNP.
2. Fluka Particle Transport Code. URL: https://ru.wikipedia.org/wiki/FLUKA.
3. Penelope. A Code System for Monte Carlo Simulation of Electron and Photon Transport URL: http://www.mcnpvised.com/visedtraining/penelope/penelope0.pdf.
4. Lessons and Training Examples on Geant4. URL: https://dev.asifmoda.com/geant4.
5. Cinelli G., Tositti L., Mostacci D., Bare J. Calibration with MCNP of NaI Detector for the Determination of Natural Radioactivity Levels in the Field. Journal of Environmental Radioactivity. 2019;155;156:31-37.
6. Mouhti I., Elanique A., Messous M.Y. Monte Carlo Modelling of a NaI(Tl) Scintillator Detectors Using MCNP Simulation Code. J. Mater. Environ. Sci. 2017;8;12:4560-4565.
7. Bagayev K.A., Kozlovskiy S.S., Novikov I.E. Program for 3D Simulation Modeling of Detection and Registration Systems of Ionizing Radiation Based on a Developed Graphical Interface. ANRI. 2007;4:35-40 (In Russ.).
8. Spectrometers-Radiometers of Gamma, Beta and Alpha Radiation MKGB-01 “RADEK”: Operation Manual. St. Petersburg, Nauchno Tekhnicheskiy Tsentr Radek Publ., 2012. 60 p. (In Russ.).
9. Scintillation Detectors of Ionizing Radiation Based on Sodium Iodide Crystals Activated by Thallium. TU 2651-001-26083472-2015. Usolye-Sibirskoye, Kristall, 2015. 10 p. (In Russ.).
10. Kapitonov M.I. Yadernaya Rezonansnaya Fluorestsentsiya = Nuclear Resonance Fluorescence.Textbook. Moscow, MGU im. M.V.Lomonosova Publ., 2018. 128 p. (In Russ.).
11. Aref’yeva D.V., Firsanov V.B., Kuruch D.D., et al. Calibration of a Gamma-Ray Scintillation Spectrometer Using the Mathematical Modeling Method. Radiatsionnaya Gigiyena = Radiation Hygiene. 2020;13;4:93-100 (In Russ.). doi: 10.21514/1998-426X-2020-13-4-93-100. EDN ZAAYGU..
12. Silant’yev A.N. Spektrometricheskiy Analiz Radioaktivnykh Prob Vneshney Sredy = Spectrometric Analysis of Radioactive Samples of the External Environment. Leningrad, Gidrometeorologicheskoye Izdatel’stvo Publ., 1969. 185 p. (In Russ.).
13. Malysheva T.A. Chislennyye Metody i Komp’yuternoye Modelirovaniye. Laboratornyy Praktikum po Approksimatsii Funktsiy. Tutorial. St. Petersburg, ITMO Publ., 2016. 33 p. (In Russ.).
PDF (RUS) Full-text article (in Russian)
Conflict of interest. The authors declare no conflict of interest.
Financing. The study had no sponsorship.
Contribution. Article was prepared with equal participation of the authors.
Article received: 20.02.2025. Accepted for publication: 25.03.2025.