Medical Radiology and Radiation Safety. 2026. Vol. 71. № 6
DOI:10.33266/1024-6177-2026-71-6-156-164
R. Bouchouareb1, K. Hadj Amor1, S. Berhab2
PARAMETRIC STUDY OF AN ANTENNA DETECTING BREAST CANCER
1 Department of Networks and Telecommunications, Oum El Bouaghi University, Institute of Technology, Oum El Bouaghi, Algeria
2 Higher National School of Telecommunication and Information and Communication Technologies, Si Radouane City, Oran, Algeria
Contact person: Rachida Bouchouareb, e-mail: This email address is being protected from spambots. You need JavaScript enabled to view it.
ABSTRACT
Relevance: Cancer diagnosis typically involves a comprehensive medical evaluation, including a physical examination, laboratory tests, and imaging techniques such as X-rays, ultrasound, CT scans, magnetic resonance imaging (MRI), and mammography. Biopsies are also performed to examine suspicious tissue samples. However, these traditional diagnostic methods have several limitations.
Purpose: Conventional breast imaging techniques suffer from limitations related to radiation exposure, discomfort, limited sensitivity in dense tissue, or high cost. This study investigates a non-ionizing, antenna-based microwave detection system as a preliminary mass-scree-
ning tool, focusing on its ability to identify dielectric inhomogeneities associated with malignant breast tumors and to differentiate normal from potentially pathological tissue.
Material and methods: Among new methods, patch antennas are a promising solution. In this study, we propose a patch antenna system for breast cancer detection. The parameters investigated using High-Frequency Structure Simulator (HFSS) software include the distance between the antenna and the breast, tumor size, tumor location, and tumor number.
Results: Simulation results of the proposed antenna for several cases are presented and analyzed. The results show that optimal reflectivity (S11) and voltage standing wave ratio (VSWR) characteristics are achieved with a distance of 10 cm between the transmitting (TX) and receiving (RX) antennas. Regarding tumor location, the highest S11 value is achieved when the tumor is located in the center of the breast. For tumor thickness, six cases were studied, revealing that larger tumors yield higher S11 values. Furthermore, the number of tumors also affects the S11 parameter, increasing as the tumor number decreases.
Conclusion: This study demonstrates the influence of tumor presence, position, size, and number on key antenna parameters (S11, S21, and VSWR), highlighting the potential of microwave antenna systems for breast cancer detection.
Keywords: patch antenna, medical antenna, breast cancer, HFSS, S11, VSWR
For citation: R. Bouchouareb, K. Hadj Amor, S. Berhab. Parametric Study of an Antenna Detecting Breast Cancer. Medical Radiology and Radiation Safety. 2026;71(6):156–164. DOI:10.33266/1024-6177-2026-71-6-156-164
References
1.Umar. A, Atabo. S. A. review of imaging techniques in scientific research/clinical diagnosis. // MOJ Anat Physiol, 2019, vol. 6, no 5, p. 175-83. DOI: 10.15406/mojap.2019.06.00269
2.Hussain, S., Mubeen, I., Ullah, N., Shah, S. S. U. D., Khan, B. A., Zahoor, M.,. & Sultan, M. A. (2022). Modern diagnostic imaging technique applications and risk factors in the medical field: a review.// BioMed research international, 2022(1), 5164970. PMID: 35707373 PMCID: PMC9192206 DOI: 10.1155/2022/5164970
3.Rana SP, Dey M, Tiberi G, Sani L, Vispa A, Raspa G, Duranti M, Ghavami M, Dudley S. Machine Learning Approaches for Automated Lesion Detection in Microwave Breast Imaging Clinical Data. Scientific Reports. 2019;9:10510. doi:10.1038/s41598-019-46974-3
4.Khan, A. Q., Touseeq, M., Rehman, S., Tahir, M., Ashfaq, M., Jaffar, E., & Abbasi, S. F. (2025). Advances in breast cancer diagnosis: A comprehensive review of imaging, biosensors, and emerging wearable technologies. Frontiers in Oncology, 15, 1587517. https://doi.org/10.3389/fonc.2025.1587517.
5.Sasada, S., Oda, N., Fukuda, H., Fujita, H., Okano, Y., Kondoh, T., et al. (2020). Microwave breast imaging using rotational bistatic impulse radar for the detection of breast cancer: Protocol for a prospective diagnostic study. JMIR Research Protocols, 9(10), e17524. https://doi.org/10.2196/17524
6.Shadrack O. A, John A. H, Graham. B, Yang. W. Portable noninvasive technologies for early breast cancer detection: A systematic review. Computers in Biology and Medicine, Volume 182, November 2024, 109219. https://doi.org/10.1016/j.compbiomed.2024.109219
7.Chishti, A. R., Aziz, A., Qureshi, M. A et al. Advances in antenna-based techniques for detection and monitoring of critical chronic diseases: A comprehensive review. //IEEE Access, (2023). 11, 104463-104484. DOI 10.1109/ACCESS.2023.3316149
8.Mahfuz, M. H., Mahfouz, A. M., Baladi, E., et al. Microstrip Patch Antennas for Breast Tumor/Cancer Cell Detection–Challenges, Designs and Future Opportunities: A Review.// IEEE Access. (2024). DOI 0.1109/ACCESS.2024.3496117
9.Saravanan, S., Haimavathi, K. U., & Aafizaa, K. Biomedical Antenna Design: Micro strip Patch Antennas for Wearable and Implantable Healthcare Applications—A Review.// Biomedical Materials & Devices, (2025). 1-11. DOI: 10.1007/s44174-025-00299-0
10.Govind. A, Paramjot. M, Ameya. S, et al. Systemic Overview of Microstrip Patch Antenna’s for Different Biomedical Applications. // National library of Medicine, Adv Pharm Bull. 2020 Jul 1;11(3):439–449. doi: 10.34172/apb.2021.051
11.Guetaf, B. Design of a New Antenna for Biomedical Applications. // Doctoral dissertation. (2024). University of 8 Mai 1945 Guelma. Algeria
12.Wang, L. Microwave imaging and sensing techniques for breast cancer detection. // Micromachines, (2023). 14(7), 1462. https://doi.org/10.3390/mi14071462.
13.El Misilmani, H. M., Naous, T., Al Khatib, S. K., et al. A survey on antenna designs for breast cancer detection using microwave imaging.// IEEE Access, (2020). 8, 102570-102594. DOI: 10.1109/ACCESS.2020.2999053
14.Wilkinson, L., & Gathani, T. Understanding breast cancer as a global health concern. // The British journal of radiology, (2022). 95(1130), 20211033. DOI: 10.1259/bjr.20211033
15. Houghton, S. C., & Hankinson, S. E. Cancer progress and priorities: breast cancer. Cancer epidemiology.// biomarkers & prevention, (2021). 30(5), 822-844. DOI: 10.1158/1055-9965.EPI-20-1193
16.Haider, R. Anatomy of the Breast. //International Journal of Scientific Multidisciplinary Research, (2023). 1(5), 401-22. https://doi.org/10.55927/ijsmr.v1i5.4394
17.Zoya, G, Alethea, R, David, W, Jacki, L, M, et al. Maternal Factors, Breast Anatomy, and Milk Production During Established Lactation—An Ultrasound Investigation.// J Imaging. (2025). Sep 12;11(9):313. doi: 10.3390/jimaging11090313
18.Biswas, S. K., Banerjee, S., Baker, G. W, et al. The mammary gland: basic structure and molecular signaling during development.// International Journal of Molecular Sciences, (2022). 23(7), 3883. PMID: 35409243 PMCID: PMC8998991 DOI: 10.3390/ijms23073883
19.Schenkel, J. M., & Pauken, K. E. Localization, tissue biology and T cell state implications for cancer immunotherapy. // Nature Reviews Immunology, (2023). 23(12), 807-823. PMID: 37253877 PMCID: PMC11448857 DOI: 10.1038/s41577-023-00884-8
20.Mohcene. A. (2022). Study and Design of a New Antenna.// Master, University May 8, 1945 Guelma,. Algeria
21.Periferakis, A., Periferakis, K., Iftime, A., et al. Detection of Acupuncture Points and Meridians Based on their Electrical Properties: Current Evidence and Future Research Perspectives. // Rom. J. Mil. Med, (2025). 128, 378-400. DOI: 10.55453/rjmm.2025.128.5.2
22/Jia, B., Huang, H., Dong, Z., Ren, X., et al. Degradable biomedical elastomers: paving the future of tissue repair and regenerative medicine. //Chemical Society Reviews, (2024). 53(8), 4086-4153. https://doi.org/10.1039/D3CS00923H
23.Fan, W., Hou, S., & Zhou, Z. The Duck 1000 Genomes Project: Achievements and perspectives.// Animal Research and One Health, (2024). 2(4), 366-376. https://doi.org/10.1002/aro2.89
24.Muñoz Taboada, G., Dahis, D., Dosta, P., et al. Sprayable hydrogel sealant for gastrointestinal wound shielding.// Advanced Materials, (2024). 36(24), 2311798. https://doi.org/10.1002/adma.202311798
25.Thomson, R. E., & Emery, W. J. Data analysis methods in physical oceanography.// Edition 4. Elsevier. (2024).
26.Ahmad. A. A, Allan. M. A, Mohd. N. M. Y, et al. Existing and Emerging Breast Cancer Detection Technologies and Its Challenges: A Review.// Appl. Sci. 2021, 11(22), 10753; https://doi.org/10.3390/app112210753.
27.Kenneth R.F, Herman P.S, Dielectric properties of tissues and biological materials: acritical review, department of bioengineering university of Pennsylvania, volume 17 Issue 1 1989
28.Coey. J. M. D, Stuart S.P. P. Handbook of Magnetism and Magnetic Materials. // Springer First Online, 02 September 2021
29.Tsabeeh. S. M. M, Adnan. M, Salah. M. M, et al. Enhancing Multispectral Breast Imaging Quality through Frame Accumulation and Hybrid GA-CPSO Registration.// Bioengineering 2024, 11(12), 1281; https://doi.org/10.3390/bioengineering11121281
30.Tale. S, Sarmad. N. M. Sahar, S. et al. Ultra-Wideband (UWB) Antennas for Breast Cancer Detection with Microwave Imaging: A Review,// Results in Engineering, (2025) vol. 25, no. 4, 104167, Mar. 2025, DOI: 10.1016/j.rineng.2025.104167.
31.Şahin. Y, Muhammed. B. K. Breast Cancer Detection Using a High-Performance Ultra-Wideband Vivaldi Antenna in a Radar-Based Microwave Breast Cancer Imaging Technique,// Applied Sciences, (2025) vol. 15, no. 11, 6015, 2025, DOI: 10.3390/app15116015.
32.Islam. M. T., Mahmud. M. Z., Islam, M. T et al, A Low-Cost and Portable Microwave Imaging System for Breast Tumor Detection Using UWB Directional Antenna Array.// Scientific Reports, (2019) vol. 9, 51620, 2019, DOI: 10.1038/s41598-019-51620-z.
33.Mariya. L, Dijana. P, Leah. M et al. A Large-Scale Study of the Ultra-Wide Band Microwave Dielectric Properties of Normal, Benign and Malignant Breast Tissues Obtained from Cancer Surgeries.// Physics in Medicine and Biology, (2007) vol. 52, pp. 6093–6115, 2007, DOI: 10.1088/0031-9155/52/20/002.
34.Yasmine. E. H, Mohamed. B, Abdelouahab. E.H et al. Parametric Study and Design of Ultra-Wideband Planar Microstrip Antenna for Breast Tumor Detection. //International Journal of Advanced Research (IJAR), 2016, DOI: 10.21474/IJAR01/3066.
35.Origlia, C., Rodriguez-Duarte, D. O., Tobon Vasquez, J. A et al. Review of microwave near-field sensing and imaging devices in medical applications. Sensors, (2024). 24(14), 4515. https://doi.org/10.3390/s24144515
36.Cheng, Y., & Fu, M. (2018). Dielectric properties for non‐invasive detection of normal, benign, and malignant breast tissues using microwave theories. Thoracic Cancer, 9(4), 459–465. https://doi.org/10.1111/1759-7714.12605
37.Bah, M. H., Hong, J. S., & Jamro, D. A. (2015). Study of breast tissues dielectric properties in UWB range for microwave breast cancer imaging. In International Conference on Computer Information Systems and Industrial Applications (CISIA 2015) (pp. 473–475). DOI 10.2991/cisia-15.2015.129
38.Deepshikha. B, Phadungsak. R and Krit J. Microwave Imaging for Breast Cancer Detection- A Comprehensive Review.// Engineered Science Publisher LLC 2024 DOI: https://dx.doi.org/10.30919/es1116
39.Jinying. W, Shuical. W, Zeyi. W. Influences of blood flow parameters on temperature distribution during liver tumor microwave ablation.// Front. Biosci. (Landmark Ed) 2021, 26(9), 504–516; https://doi.org/10.52586/4963
40.Koutar. H. A, Ayoub. M. Creation, Simulation, Realization and Analysis of a Planar Antenna (Microstrip Patch Antenna).// License, Oum el baoughi, Larbi Ben M’Hidi University, 2022.
41.Zahra. K. Développement d’antennes pour la détection de tumeurs cancéreuses.// Liban, Doctoral dissertatio Universite libanaise, 2018
42.Fawzia. A. A, Askin. D. A novel textile‑based UWB patch antenna for breast cancer imaging.// Physical and Engineering Sciences in Medicine (2024) 47:851–861. https://doi.org/10.1007/s13246-024-01409-w
43.Elsheikh, D. and Eldamak, A.R. Microwave Textile Sensors for Breast Cancer Detection.// 38th National Radio Science Conference (NRSC), Mansoura, 27-29 July 2021, 288-294. https://doi.org/10.1109/NRSC52299.2021.9509829
44.Xiaoyou. L, Yifan. C, Zheng, G et al. Ultra-Wideband Textile Antenna for Wearable Microwave Medical Imaging Applications.//. February 2020 IEEE Transactions on Antennas and Propagation PP(99):1-1. DOI: 10.1109/TAP.2020.2970072
45.Fernández-Aranzamendi, E. G., Castillo-Araníbar, P. R., San Román Castillo, E. G., Oller, B. S., Ventura-Zaa, L., Eguiluz-Rodriguez, G., González-Posadas, V., & Segovia-Vargas, D. (2024). Dielectric characterization of ex-vivo breast tissues: Differentiation of tumor types through permittivity measurements. Cancers, 16(4), 793. https://doi.org/10.3390/cancers16040793.
46.Lazebnik et al. measured significant dielectric property differences across breast tissue types in the microwave band, forming a foundational basis for imaging contrast (e.g., malignant vs. normal tissue) DOI: 10.3389/fphys.2017.00533 (review citing Lazebnik et al.) MDPI
47.Lazebnik, M., Zhu, C., Palmer, G. M., Harter, J., Sewall, S., Ramanujam, N., & Hagness, S. C. (2008). Electromagnetic spectroscopy of normal breast tissue specimens obtained from reduction surgeries: Comparison of optical and microwave properties. IEEE Transactions on Biomedical Engineering, 55(10), 2444–2451. https://doi.org/10.1109/TBME.2008.925700
48.Garrett, J. D., & Fear, E. C. (2015). Average dielectric property analysis of complex breast tissue with microwave transmission measurements. Sensors, 15(1), 1199–1216. https://doi.org/10.3390/s150101199
49.Aleksandar. J, Ibrahim. A, Mehmet. C, Onur. B, Mustafa .E .A. Microwave Breast Lesion Classification – Results from Clinical Investigation of the SAFE Microwave Breast Cancer System. Academic Radiology, Volume 30, Supplement 2, September 2023, Pages S1-S8. https://doi.org/10.1016/j.acra.2022.12.001
50.Shea, J. D., Kosmas, P., Hagness, S. C., & Van Veen, B. D. (2010). Three-dimensional microwave imaging of realistic numerical breast phantoms via a multiple-frequency inverse scattering technique. Medical Physics, 37(8), 4210–4226. https://doi.org/10.1118/1.3443569.
51.Särestöniemi, M., Singh, D., Dessai, R., Heredia, C., Myllymäki, S., & Myllylä, T. (2024). Realistic 3D phantoms for validation of microwave sensing in health monitoring applications. Sensors, 24(6), 1975. https://doi.org/10.3390/s24061975.
52.Allin. J. D, Gopika. S, Sobiya, S. M et al . Breast Cancer Detection Using Patch Antenna.// 2nd International Conference on Advancements in Electrical, Electronics, Communication, Computing and Automation (ICAECA) 2023, DOI: 10.1109/ICAECA56562.2023.10200909.
PDF (RUS) Full-text article (in Russian)
Conflict of interest.The authors declare no conflict of interest.
Financing. The study had no sponsorship.
Contribution. Article was prepared with equal participation of the authors.
Article received: 20.06.2026. Accepted for publication: 25.07.2026.